Set the Line 10 inline alcohol meter (lab adjustment)
Adjust the Line 10 inline alcohol reading to match the QA lab result using the Laboratory Adjustment screen.
Adjust the Line 10 inline alcohol reading to match the QA lab result using the Laboratory Adjustment screen.
1
Step 1 of 6
If the inline Anton Paar is continually out compared to Lab results the measurement can be adjusted.
Select Laboratory Adjustment, then select Login.
Before changing ensure Lab results are accurate and have been verified.
2
Step 2 of 6
Select Administrator as the user name and enter the password: administrator. Select OK.
3
Step 3 of 6
Ensure the filler is running.
On the Laboratory Adjustment screen, check that the 'A' block shows under Reference. If it does not, press Save values.
4
Step 4 of 6
Make sure a can has been checked by QA and you have the results.
Select 'A' and enter the QA result.
5
Step 5 of 6
Enter the value given by Lab on the keypad and select OK.
6
Step 6 of 6
Select Finish adjustment.
After 10 minutes, verify the alcohol reading is still in line with the lab result.