ULTRAMETER pH & ORP calibration
1
Step 1 of 10
Rinse cell cup 3 times with 7.0 buffer solution.
2
Step 2 of 10
Fill cell cup with 7.0 buffer solution
3
Step 3 of 10
Press pH, then CAL. "CAL", "7" and "BUFFER" will appear on the screen
4
Step 4 of 10
Press "up" or "down" butom until display agrees with buffer value
5
Step 5 of 10
Press CAL once to accept the value
6
Step 6 of 10
Rinse 3 times with acid (1-6) buffer solution
7
Step 7 of 10Repeat step 4 and 5
8
Step 8 of 10
Rinse 3 times with base ( 8-14) buffer solution
9
Step 9 of 10Repeat step 4 and 5
10
Step 10 of 10
Press CAL to exit