ULTRAMETER pH & ORP calibration
1
Step 1 of 10
<p>Rinse cell cup 3 times with 7.0 buffer solution.</p>

Rinse cell cup 3 times with 7.0 buffer solution.

2
Step 2 of 10
<p>Fill cell cup with 7.0 buffer solution</p>

Fill cell cup with 7.0 buffer solution

3
Step 3 of 10
<p>Press pH, then CAL. &quot;CAL&quot;, &quot;7&quot; and &quot;BUFFER&quot; will appear on the screen</p>

Press pH, then CAL. "CAL", "7" and "BUFFER" will appear on the screen

4
Step 4 of 10
<p>Press &quot;up&quot; or &quot;down&quot; butom until display agrees with buffer value</p>

Press "up" or "down" butom until display agrees with buffer value

5
Step 5 of 10
<p>Press CAL once to accept the value</p>

Press CAL once to accept the value

6
Step 6 of 10
<p>Rinse 3 times with acid (1-6) buffer solution</p>

Rinse 3 times with acid (1-6) buffer solution

7
Step 7 of 10

Repeat step 4 and 5

8
Step 8 of 10
<p>Rinse 3 times with base ( 8-14) buffer solution</p>

Rinse 3 times with base ( 8-14) buffer solution

9
Step 9 of 10

Repeat step 4 and 5

10
Step 10 of 10
<p>Press CAL to exit</p>

Press CAL to exit